See: Description
| Interface | Description |
|---|---|
| Hardware |
Provides low-level hardware operations.
|
| Transport |
Represents a communications link over which commands are sent to the device
and replies received from it.
|
| Class | Description |
|---|---|
| ASCIIModbus |
Very simple ASCII Modbus communications, just good enough for the Lighthouse particle counter.
|
| ChannelDatum |
Represents a single channel datum from a Lighthouse counter.
|
| Client |
Exports high-level requests made of a Lighthouse particle counter.
|
| DataRecord |
Represents a single data record from a Lighthouse particle counter.
|
| FTDITransport |
Sends and receives bytes using an FTDI USB-to-serial cable.
|
| LighthouseDriver |
Shell command driver.
|
| NetTransport |
Sends and receives bytes using a serial port on the local host.
|
| SerialTransport |
Sends and receives bytes using a serial port on the local host.
|
| Enum | Description |
|---|---|
| ChannelType |
The kinds of things Lighthouse data channel might measure.
|
| ChannelUnits |
The measurement units for Lighthouse data channels.
|
| DataChannel |
The data channels we can read from a Lighthouse counter.
|
| DeviceCommand |
The hardware commands used by this software.
|
| DeviceFlag |
The status flags that can be set for a Lighthouse counter device.
|
| RecordFlag |
The device status flags that can be set for a Lighthouse data record.
|
The device measures the density of airborne particles of various sizes. The models purchased for the LSST I&T clean room have attachments for the measurement of temperature and relative humidity as well.
The class LighthouseDriver can be run as a CCS command-line
application using the script run-counter.sh. The class Client
exports the highest-level operations, such as fetching records. The Hardware interface represents
data items which map directly to one or more registers in the device. It's implemented by ASCIIModbus
which generates command frames and interprets response frames according to the ASCII Modbus protocol.
The Transport interface represents the means by which frames are transmitted to and from the device.
There are three implementations of Transport which use the package org.lsst.ccs.drivers.ascii:
SerialTransport, FTDITransport and NetTransport.
The device comes with an FTDI USB-to-serial conversion
cable that's USB-A on one end and RJ-45 on the other. Most modern Linux installations have
the ftdi_sio and usbserial kernel modules enabled, so that when the device cable is plugged
in to a USB port the modules are loaded if needed and the device is treated as a serial
TTY with a name such as /dev/ttyUSB0. You'll need to check the output of the dmesg command
after you plug in the cable to see exactly what name has been assigned. The SerialTransport
class and the driver's openSerial command are used to communicate with a Lighthouse device installed
in this way. The comm parameters for the serial port are fixed: raw tty mode, 19200 bps, eight
data bits, no parity, one stop bit. This is the factory setting for the device.
It's also possible to communicate with the FTDI chip directly. For this you have
to make sure that ftdi_sio and usbserial are never loaded on your Linux system. Each FTDI chip has a
unique serial number which you can use to identify the device you
want, removing the need to discover dynamic device name
assignments. The FTDI driver page
tells how to set this up
on a RedHat-like Linux system. There's a server program org.lsst.ccs.drivers.ftdi.FtdiServer
allowing remote access to a host's locally installed FTDI devices; it listens on TCP port 9001. In
this case the remote user must precede the FTDI serial number with the server's host name (or IP
address) followed by a colon. The Lighthouse driver FTDITransport class and the ftdiOpen command
implement the raw-FTDI mode of operation.
Lastly, the NetTransport class and the driver's netOpen command allow you to forward the ASCII Modbus frames to an arbitrary TCP server at a specified host and port, and to collect responses from that server. The server presumably communicates with the actual Lighthouse device somehow and acts as a relay.
Whatever transport variant is used, each device is assumed to be attached to its host using its own cable and to have a Modbus slave ID of 1 (the factory setting).
Despite the settings of the built-in display, the particle counts read from the device are always raw, cumulative counts. For example, a value of 2259 for the 1.0-micron counter channel indicates that during the sample period the device counted 2259 particles of size 1.0 microns or larger.
Although the display may show temperature in degrees Celsius or Fahrenheit, temperatures read from the device are always in degrees Fahrenheit.